Note: You do not need any special tools to complete the adjustment for the Gilson Pipetman M.
Recommend Supplies:
- Balance
- Pipette Tips
Step 1:
Make sure the pipette has enough battery life before proceeding to the next step.

Step 2:
Start by turning on the pipette by pressing down on the dispense button.

Step 3:
Prior to adjusting the pipette, collect 2-3 data points at each test volume (low, middle, and high) on the balance. The data will be used in Step 7.

Step 4:
After the data points have been collected, click on the blue triangle button.

Step 5:
Turn the adjustment wheel counter-clockwise to “Configuration” and press down on the dispense button.

Step 6:
Turn the adjustment wheel counter clock-wise to “Adjustment” and press down on the dispense button.

Step 7:
The standard readjustment utilizes a 3-point calibration (low, middle & high). Using the data collected in step 1, get an average value in microliters for each test volume.

Step 8:
Click the dispense button (the current setpoint value will flash) to modify the low point value and adjust to the calculated average value by turning the adjustment knob. Once the value is set, click the dispense button to lock in the value.

Step 9:
To move to the middle & high test points, turn the adjustment knob clockwise. Repeat the same process in step 8 for the middle and high averaged values.

Step 10:
After all values are entered, turn the adjustment knob clockwise to “Click to confirm”. Click the dispense button. This will cause the pipette to restart.

Note: There is an option to perform a factory reset of the calibrated values. To begin, complete steps 1 through 6. After completing step 6, turn the adjustment knob clockwise to select “Reset settings.” The prompt “Click to confirm” will appear on the screen. Press the dispense button to confirm and complete the factory reset.

Step 11:
Use the balance to check the pipette now that the adjustment has been made.
Related Posts:
Gilson Pipetman M & VWR Electronic – Battery Replacement
What model of Gilson Pipette do I have?
Styles of Gilson Pipette Tip Ejectors
Disclaimer: Any action you take using the information on this website is strictly at your own risk. The information herein does not constitute professional advice and is general in nature. We make no warranty that this information will meet your requirements, be safe, accurate, or error-free. Pipette Supplies, Inc. is not responsible for any errors or omissions, any results obtained from the use of this information, or any loss or damage arising out of the use of this information. This site is for educational purposes only.
Fair Use: Copyright Disclaimer under section 107 of the Copyright Act of 1976, allows for “fair use” for purposes such as comments, criticism, teaching, scholarship, news reporting, and research.
Fair use is permitted by copyright statute that might otherwise be infringing.
Leave a Reply